Application
OTR-1903 is based on the coulometric sensor method and available for the OTR test of flexible films, sheeting and packages.

Films:
Plastic films, aluminum foil, aluminized films, plastic composite films, paper-plastic composite films, coextruded films, aluminum foil composite films and many others.

Packages:
Plastic, rubber, paper, paper-plastic composite, glass and metal packages, e.g. red wine bottles, blister, various plastic bottles, Coke bottles, peanut oil packages, Tetra Pak materials, vacuum bags, metal three-piece cans, plastic packages for cosmetic, soft tubes for toothpaste, jelly and yogurt cups.

Principles

Film:
The specimen is clamped between the upper and lower chamber, the upper chamber is purged by a oxygen stream, and the other side of specimen, lower chamber is purged by a stream of nitrogen, due to the effect of oxygen concentration gradient at two sides of the specimen, oxygen molecules will permeate through the specimen into the nitrogen side, penetrated oxygen molecules will be carried to the coulometric sensor which generates the
proportional signal, so oxygen transmission rate and permeability coefficient will be obtained
by calculating the sensor signal.

Package:
The package is mounted on the lower chamber, the outside of the package is air or a pure
oxygen stream, the inside of the package is purged by a nitrogen stream, oxygen will
permeate through the wall into the inside of the package, penetrated oxygen molecules will
be carried to the coulometric sensor which generates the proportional signal, so oxygen
transmission rate will be obtained by calculating the sensor signal.

Features

  • Coulometric sensor method
  • Own technology
  • Film and package test
  • one piece specimen test
  • One-button test, test automatically in whole process
  • Multiple
  • Data curves display
  • Computer control
  • Digital flow adjustment
  • Chamber with temperature sensor embedded in
  • Environment temperature and humidity sensor inside
  • Chambers with water bath temperature control technology
  • Unit based on embedded system,24bit Δ-Σ AD
  • Pressure, oxygen and temperature overload protection
  • Traceable reference film calibration

Supports DSM system (DSM, lab data management

Technical indexes

  •  Test range:
    Film: 0.01-6300 cm3/m2·d (Standard)
    0.07-6,3000 cm3/m2·d (Optional)                                                                                                                                             Resolution: 0.001 cm3/m2·d
  • Package: 0.0001~62 cm3/pkg·d
    Resolution: 0.00001 cm3/m2·d
  • Specimen amount: one piece
  • Temperature range: 5C – 95C (with temperature controller)
  • Temperature accuracy: ±0.1C
    Humidity range: 0%RH,35%RH-90%RH,100%RH
  • Humidity accuracy: ±1%RH
  • Test area: 50cm2
  • Specimen size:
    Film: size:>=150mm * 94mm;
  • thickness: less 1mm
  • Package: max size: D120mm * 400mm
  • Carrier gas: 99.999% Nitrogen
  • Carrier flow: 0~20ml/min
  • Gas interface: 1/8inch
  • Power supply: AC 220V 50Hz
  • Dimension: 330mm(L) * 600mm(W) * 330mm(H)
  • Net weight: 28kg

Standard

ASTM D3985, ASTM F2622, ASTM F1307, ASTM F1927, ISO 15105-2, GB/T 19789, JIS K7126-B, YBB 00082003

Configuration

Delivery: Main unit, computer, OX-TRANS software, tank regulator, PCIE card, communication cable, sample cutter, vacuum grease.                                                                                                                                                                      Optional: Temperature controller, package fixtures, reference film, vacuum grease, sample cutter, DSM system